Troubleshooting the stabilization process
Troubleshooting the stabilization process
The projector stabilization has stopped unexpectedly
Possible causes:
•No communication with projector. The projector may be switched off, removed, or the
•No communication with optical sensor. The sensor connection to the PC may be disconnected.
•The optical sensor is not or no longer positioned on the sensor patch.
•The user has selected another signal input or another preset for the projector.
In all of these cases, you must restore any problems in the connections (if appropriate). Then you must run MediCal Pro and select the proper Preset. This should restart the stabilization.
The PC does not boot or does not recognise the keyboard after booting.
When a USB keyboard and the optical sensor are both connected to the PC’s USB bus, there may be a problem at
A workaround is to connect the keyboard to the PS/2 connector.