SECTION 13. CR10 MEASUREMENTS

FIGURE 13.5-2. Excitation and Measurement Sequence for 4 Wire Full Bridge

 

TABLE 13.5-1. Comparison of Bridge Measurement Instructions
Instr. # Circuit

Description

Instr. # Circuit

Description

4

DC Half Bridge

The delay parameter

 

 

allows the user entered

 

 

settling time com-

 

 

pensate for capacitance

 

 

in long lead lengths. No

 

 

polarity reversal. One

 

 

single-ended measure-

 

 

ment. Measured

 

 

voltage output.

5

AC Half Bridge

Rapid reversal of

 

 

excitation polarity for

 

 

ion depolarization.

 

 

One single-ended

 

 

measurement at each

 

 

excitation polarity.

 

 

Ratiometric output.

6

4 Wire

One differential

 

Full Bridge

measurement at each

 

 

excitation polarity.

 

 

Ratiometric output.

7

3 Wire

Compensates for lead

 

Half Bridge

wire resistance,

 

 

assuming resistance is

 

 

same in both wires.

 

 

Two single-ended

 

 

measurements at each

 

 

excitation polarity.

 

 

Ratiometric output.

8

Differential

Makes a differential

 

Measurement

measurement without

 

with Excitation

reversing excitation

 

 

polarity. Used for fast

 

 

measurements on load

 

 

cells, PRTs etc.

 

 

Resolution and

 

 

common mode rejection

 

 

worse than 6 if used

 

 

with delay = 0.

 

 

Measured voltage

 

 

output.

9

6 Wire

Compensates for lead

 

Full Bridge

wire resistance. Two

 

or 4 Wire

differential measure-

 

Half Bridge

ments at each

 

 

excitation polarity.

 

 

Ratiometric output.

13-19