Manual background

Rosemount 848L

Reference Manual

00809-0100-4696, Rev AA September 2004

ADVISE_ACTIVE

The ADVISE_ACTIVE parameter displays which of the advisories is active. Only the advisory with the highest priority will be displayed. This priority is not the same as the ADVISE_PRI parameter described above. This priority is not user configurable.

ADVISE_ALM

ADVISE_ALM is an alarm indicating advisory alarms. These conditions do not have a direct impact on the process or device integrity.

Recommended Actions for PlantWeb Alarms

RECOMMENDED_ACTION

The RECOMMENDED_ACTION parameter displays a text string that will give a recommended course of action to take based on which type and which specific event of the PlantWeb alarms is active.

Table 3-1.

RB.RECOMMENDED_ATION

 

Failed/Maint/Advise

Recommended Action

Alarm Type

Active Event

Text String

 

 

 

NONE

None

No action required

 

 

 

 

Prescaler Overflow

Check the Divisor parameter of all PS function

ADVISORY

 

calls in the logic equations

NV Write Deferred

Reduce the frequency in which applications

 

 

 

write to NV Memory

 

PWA Simulate Active

Disable PWA_SIMULATE parameter in the

 

 

Resource Block

 

Secondary Value

Ensure that the transmitter is not too close to

MAINTENANCE

Degraded

extreme hot or cold environments

 

 

Configuration Error

Verify that the Logic equations are correct in

 

 

 

the Logic transducer block

 

 

 

 

Electronics Failure

Replace the electronics

 

NV Memory Failure

Replace the electronics

 

No I/O Power

Check the IO Power supply, polarity, wiring,

FAILED

 

and connections.

Primary Value Failure

Check the sensor, configuration, wiring, and

 

 

 

connection for open or shorted sensors.

 

Secondary Value

Verify that the body temperature is within the

 

Failure

operating limits of this device.

 

 

 

3-8

Page 28
Image 28
Emerson manual Rosemount 848L, Reference Manual, 00809-0100-4696,Rev AA September