EQ - Channels 17-32 are set by the result of an equation specified by up to 80 characters and stored in parameter EQx, where x is the equation number. The equation results are available as a discrete value and status in parameter EQx_VALUE. They may be referenced by the EQ (u) function. The range of ‘u’ is 1 to 16. These are intended to be intermediate values that are used because the value is used in other equations or because the equation text was too long. A configured DI block may use an equation channel (range 17 to 36) in order to make the result available to other devices.

OUT – The value will be the same as the requested output.

Additional Channels

The ten channels that are used for connecting multiple in or out function blocks can not be referred to by equation functions. Each has a status that is Bad if any Input status is Bad. Channels 37 to 41 pack the values into one byte so that a DI or DO function block can read or write them. Any block linked to a DI block with packed data must be capable of handling the packed boolean values. Channels 42 to 46 may be used with standard MDI or MDO blocks.

Channel

Type

FB

 

 

 

37

All IN

DI

 

 

 

38

All DO

DO

39

First 8 EQ

DI

 

 

 

40

Last 8 EQ

DI

41

All OUT

DO

 

 

 

42

All IN

MDI

43

All DO

MDO

 

 

 

44

First 8 EQ

MDI

45

Last 8 EQ

MDI

 

 

 

46

All Out

MDO

 

 

 

Reading a channel value will reset all of the channel latches that are configured. Channels 38 and 43 do not have latches. If a DI and a MDI are both used they will interfere with the latches, but the user is expected to use one or the other, never both.

3-14

Page 34
Image 34
Emerson manual Rosemount 848L, Reference Manual, 00809-0100-4696,Rev AA September