Emerson AC4 user manual 7.8.2Turn Standby Testing On or Off, 7.8.3Specify Failed Standby Response

Models: AC4

1 89
Download 89 pages 35.25 Kb
Page 51
Image 51
7.8.2Turn Standby Testing On or Off

System and Control Options

7.8.2Turn Standby Testing On or Off

Standby testing permits scheduling an automatic operational check of devices in Standby mode.

When this feature is activated (ON), the Liebert AC4 tests all devices in Standby mode by putting each device in operating mode for a designated time. You may specify the time of day the testing begins, the duration of the test and the interval between tests. Standby testing is OFF (deactivated) by default.

To change the standby testing settings:

From the Setup System Operation Menu, use the arrows ↑↓ to choose Standby Testing, then press Enter .

In the Change Standby Testing screen, press Enter to edit Standby Testing, then use the arrows ↑↓ to choose ON to activate or OFF to deactivate the standby testing feature. Press Enter .

Select the desired interval for how frequently the testing should occur in the Every 03 Days field. The default is 3 days; valid entries range from 0 to 99 days.

Setup Operation Menu

AUTO SEQUENCING OFF

>STANDBY TESTING OFF

FAILED STANDBY N ↑↓=NEXT =SELECT

HOLD DELAY

:

10:00

RESTART

:

00:06

For each digit, use the arrows ↑↓ to choose a number from 0 to 9, then press Enter to advance to the next digit.

When finished, press Enter to configure the time of day.

Specify the time of day the test should begin in the At 00:00 field. The format is HH:MM for hours and minutes.

Use the arrows ↑↓ to choose from available entries for hours (00-23), then press Enter to advance to minutes.

Use the arrows ↑↓ to select minutes (00-59).

When finished, press Enter again.

Change Standby Testing

>STANDBY TESTING ON EVERY 03 DAYS

AT 00:00 FOR 10 MIN ↑↓=NEXT =EDIT

Specify the duration of the test in minutes—the length of time the standby units will be turned On—in the For 10 Min field. Use the arrows ↑↓ to choose a number from 0 to 59 minutes. The default is 10 minutes.

When finished, press Enter .

7.8.3Specify Failed Standby Response

The Liebert AC4’s Failed Standby feature determines what happens when an operating device fails and a standby device then becomes an operating unit. A subsequent alarm in the newly operating unit (previously the standby device) results in either restarting the previously operating device that was alarmed or not. This feature specifies whether each device remains in its current state or returns to its previous state of operation.

When this feature is activated (Y), each device returns to its previous state, as Table 19 shows: the newly enabled Standby device is disabled and the previously operating device is returned to Operating mode. The default setting is N (deactivated), meaning that each device remains in its current state: the Standby device remains enabled and the previously operating unit remains disabled.

Table 19 Failed Standby response

Failed Standby

Previous Standby Device

Previously Operating

Setting

(alarm after being enabled)

Device

 

 

 

Activated (Y)

Disabled

Returned to Operating mode

 

 

 

Deactivated (N)

Remains enabled

Remains disabled

 

 

 

To change the Failed Standby settings:

From the Setup System Operation Menu, use the arrows ↑↓ to choose Failed Standby, then press Enter .

In the Change Failed Standby screen, press use the arrows ↑↓ to choose Y to activate or N to deactivate the feature.

Press Enter .

Setup Operation Menu

AUTO SEQUENCING OFF STANDBY TESTING OFF

>FAILED STANDBY N

↑↓=NEXT =SELECT

HOLD DELAY

:

10:00

RESTART

:

00:06

Change Failed Standby

>FAILED STANDBY N

↑↓=NEXT =END

45

Liebert® AC4

Page 51
Image 51
Emerson AC4 user manual 7.8.2Turn Standby Testing On or Off, 7.8.3Specify Failed Standby Response, Setup Operation Menu