| Electrical Safety Analyzer |
| Performing Electrical Safety Tests |
Performing an Alternative Equipment Leakage | The following outlet conditions apply when performing this |
Test | test: |
Note
The alternative equipment leakage test is available when the EN62353 standard is selected.
During the Alternative Equipment Leakage test, the voltage source is applied between
This test is not applicable for equipment with internal electrical power source. The switches in mains part shall be closed during measurement.
To perform an alternative equipment leakage test:
1.Press B.
2.Press the softkey labeled Alternative Equipment.
3.Press Tto apply the voltage and read the current in the display.
Figure 24 shows the electrical connections between the Analyzer and the DUT during an Alternative Equipment Leakage Test.
•Closed Earth
•Open Earth
Note
If there are more than five applied parts to connect to the Analyzer, see Using the 1210 Adapter later in this manual.
Performing an Alternative Applied Part Leakage Test
Note
The Alternative applied part leakage test is available when the EN62353 standard is selected.
During the Alternative Applied Part Leakage test, the test voltage is applied between
asingle function and the