Electrical Safety Analyzer

 

Simulating ECG Waveforms

Measuring Current

Note

The Analyzer can make dc only, ac only, and ac+dc

If the ECG monitor/interpreter has banana posts

current measurements up to 10 mA. To make a current

instead of snaps, use the optional universal snap

to banana adapter to connect to the Analyzer.

measurement:

To access the ECG Simulation Waveform menu shown in

1. Press the softkey labeled Leakage from the

Figure 32, press C. From this menu, a number of

Point-To-Point menu.

different waveforms are selected through F1, and the rate

2. Using Gor Hselect between ac only, dc only, or

or frequency of the waveform is selected through F2.

ac+dc measurement mode.

 

3. Insert test leads in the RED (V/Ω/A ) and BLACK

 

jacks.

 

4. Place the leads on the two points the unknown

 

current may flow and read the measurement in the

 

Analyzer’s display.

 

Simulating ECG Waveforms

The Analyzer is capable of generating various waveforms on the applied parts connection posts. These signals are used to test the performance characteristics of ECG monitors and ECG strip printers. See Figure 33 for proper connections between the Analyzer and an ECG monitor. For monitors using the snap style connectors, insert the BJ2ECG adapter into the connectors at the top of the Analyzer and connect the monitor leads to the snap connectors on the adapter.

fis109.jpg

Figure 32. ECG Waveform Simulation Menu

To select one of the predefined waveforms, press the softkey labeled Wave Form. A scroll box with Knext to it appears above the softkey label. Use Gor Hto scroll through the different waveforms.

For all waveforms except VFIB and Triangle, the rate or frequency of the waveform is adjusted through the softkey

51

Page 65
Image 65
Fluke ESA612 specifications Simulating ECG Waveforms, Measuring Current