ESA612

Users Manual

 

MAINS

 

DUT_L1

DEVICE UNDER TEST

 

ON

 

 

L1

 

 

 

APPLIED

 

 

 

 

 

 

 

 

PART

MAINS

 

L2

 

 

 

 

DUT_L2

 

 

 

OPEN

 

 

 

 

 

L2

 

 

 

 

 

 

EARTH

 

 

 

 

OPEN

DUT_PE

 

 

 

 

FE

 

 

 

 

 

 

 

 

CONDUCTIVE PART

 

 

MD

 

TEST LEAD

 

PE

APPLIED PART

 

 

 

 

 

 

 

 

OPEN

 

 

 

 

 

 

faw28.eps

Figure 19. Enclosure Leakage Current Test Schematic

Note

Chassis leakage is the same schematic without the Applied Parts switch.

30

Page 44
Image 44
Fluke ESA612 specifications Enclosure Leakage Current Test Schematic