Making Point-To-Point Measurements

3. Place the probe tips across the unknown voltage and

The Analyzer can make voltage, resistance, and low

read the measurement in the Analyzer’s display.

The Analyzer will measure up to 300 volts ac.

current measurements through its Point-to-Point function.

To access the Point-to-Point function menu shown in

Measuring Resistance

Figure 31, press R. Softkeys F1 through F3 are used

To make a resistance measurement:

to select the measurement function.

1. Press the softkey labeled Resistance from the

 

 

Point-To-Point menu.

 

2. Insert test leads in the RED (V/Ω/A ) and BLACK

 

jacks.

 

3. Null lead resistance by shorting the leads together

 

and pressing the softkey labeled Zero Leads.

 

4. Place the probes across the unknown resistance and

 

read the measurement in the Analyzer’s display.

 

The Analyzer will measure resistances up to 2.0 Ω.

fis108.jpg

 

Figure 31. Point-To-Point Function Menu

 

Measuring Voltage

 

To make a voltage measurement:

 

1. Press the softkey labeled Voltage from the

 

Point-To-Point menu.

 

2. Insert test leads in the RED (V/Ω/A ) and BLACK

 

jacks.

 

50