ESA612

Users Manual

Detailed Specifications

Test Standard Selections

ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551

Voltage

 

Ranges (Mains voltage)

90.0 to 132.0 V ac rms

 

180.0 to 264.0 V ac rms

Range (Point-to-point voltage)

0.0 to 300.0 V ac rms

Accuracy

±(2 % of reading + 0.2 V)

Earth Resistance

 

Modes

Two terminal

Test Current

>200 mA ac

Range

0.000 to 2.000 Ω

Accuracy

±(2 % of reading + 0.015 Ω)

Equipment Current

 

Range

0.0 to 20.0 A ac rms

Accuracy

±(5 % of reading + (2 counts or 0.2A, whichever is greater))

Duty cycle

15 A to 20 A, 5 min. on/5 min. off

 

10 A to 15 A, 7 min. on/3 min. off

 

0 A to 10 A, continuous

Leakage Current

 

Modes*

AC+DC (True-rms)

 

AC only

 

DC only

 

* Modes: AC+DC, AC only, and DC only available for all leakages with exception of MAP

 

that are available in True-rms (shown as AC+DC)

Patient Load Selection

AAMI ES1-1993 Fig. 1

 

IEC 60601: Fig. 15

Crest factor

≤3

62

Page 76
Image 76
Fluke ESA612 specifications Detailed Specifications, Voltage