Electrical Safety Analyzer

Intended Use

Inspect the Analyzer before using it. Do not use the Analyzer if abnormal conditions of any sort are noted (such as a faulty display, broken case, etc.)

Inspect the test leads for damaged insulation or exposed metal. Check test lead continuity. Replace damaged leads before using the Analyzer.

When testing, always be sure to keep your fingers behind the safety barriers on the test leads.

Never open the Analyzer's case. Dangerous voltages are present. There are no user replaceable parts in the Analyzer.

Have the Analyzer serviced only by qualified personnel.

The Analyzer must be properly earthed. Only use a supply socket that has a protective earth contact. If there is any doubt as to the effectiveness of the supply socket earth, do not connect the Analyzer. Do not use a two-conductor adapter or extension cord; this will break the protective ground connection.

Do not use the 15-20 A adapter to power devices rated in excess of 15 A. Doing so may overload the installation.

Use extreme caution when working with voltages above 30 V.

Use the proper terminals, functions and ranges for the test being performed.

Do not touch metal parts of the device under test (DUT) during analysis. The DUT should be considered an electrical shock hazard when connected to the Analyzer as some tests involve high voltages, high currents, and/or the removal of DUT earth bond.

Intended Use

The Analyzer is intended for use by trained service technicians to perform periodic inspections on a wide range of medical equipment. The testing procedures are menu-driven, and simple to operate.

3

Page 17
Image 17
Fluke ESA612 specifications Intended Use, Electrical Safety Analyzer