Electrical Safety Analyzer Instrument Familiarization
5
Table 2. Top-Panel Controls and Connections
Item Name Description
1
Equipment Outlet
Configuration
Buttons
Controls the configuration of
the equipment outlet. Opens
and closes the neutral and
ground connection and
reverses the polarity of the
neutral and hot connection.
2 High Voltage
Indicator
Indicates when high voltage
is applied to the
ECG/Applied Parts posts or
L1 and L2 of the Test
Receptacle.
3 Test Function
Buttons
Selects the various Analyzer
test functions.
4 Navigation Buttons Cursor control buttons for
navigating menus and lists.
5 Test Button Initiates selected tests.
6 Save Button Saves the measurement or
ECG waveform to memory.
7 Input Jacks Test lead connectors.
Item Name Description
8 Nulling Jack Connection for zeroing test
lead resistance.
9 Function Softkeys
Keys F1 through F4 are used
to select from a number of
selections that appear in the
LCD display above each
function softkey.