Electrical Safety Analyzer

 

Detailed Specifications

Ranges

0.0 to 199.9 μA

 

200 to 1999 μA

 

2.00 to 10.00 mA

Accuracy

 

DC to 1 kHz

±(1 % of reading + (1 μA or 1 LSD, whichever is greater))

1 to 100 kHz

±(2 % of reading + (1 μA or 1 LSD, whichever is greater))

100 kHz to 1 MHz

±(5 % of reading + (1 μA or 1 LSD, whichever is greater))

Mains on applied part test voltage

100 % ±5 % of Mains for AAMI, current limited to 1 mA ±25 % per AAMI

 

100 % ±5 % of Mains for 62353 current limited to 3.5 mA ±25 % per 62353

 

100 % ±5 % of Mains for 60601 current limited to 7.5 mA ±25 % per 60601

Differential leakage

 

Ranges

10 to 199 μA

 

200 to 2000 μA

 

2.00 to 20.00 mA

Accuracy

±(10 % of reading + (2 counts or 20 μA, whichever is greater))

Insulation resistance

 

Ranges

0.5 to 20.0 MΩ

 

20.0 to 100.0 MΩ

Accuracy

 

20 MΩ Range

±(2 % of reading + 0.2 MΩ)

100 MΩ Range

±(7.5 % of reading + 0.2 MΩ)

Source test voltage

500 V dc (+20 %, -0 %) 1.5 ±0.25 mA short-circuit current or 250 V dc selectable

Maximum load capacitance

1 μF