Performing a Differential Leakage Current Test

Note

Note

If there are more than five applied parts to

connect to the Analyzer, see Using the 1210

The Differential Leakage Current test is available

Adapter later in this manual.

when the EN62353 standard is selected.

 

The differential leakage current test measures the magnitudes of the differential current flowing in the Equipment Outlet live and neutral, with power applied to the equipment outlet. All applied parts should be connected during this test, if equipment has applicable applied parts.

To perform a differential leakage current test:

1.Press B.

2.Press the soft key labeled Differential.

Figure 28 shows the electrical connections between the Analyzer and the DUT during a Differential Leakage Current test.

The following outlet conditions apply when performing this test:

Normal Polarity, Closed Earth

Normal Polarity, Open Earth

Reversed Polarity, Closed Earth

Reversed Polarity, Open Earth