2

1

3

4

Figure 49. J11 ConnectionsTable 23. J11 Connections

J11-1 Green/Yellow

J11-2 Black

J11-3 White

8.Connect the replacement J11 connector to the chassis. The black and white leads from the blower motor have pins installed; these pins must be inserted into the back of J11 as indicated in Figure 49.

9.Connect the ring terminals on the green/yellow wire from J11 and the green/yellow wire from the blower to the ground screw.

10.Connect the push-on terminals to capacitor C200.

11.Reinstall the blower housing cover plate and the blower cover.

120F300S/F300SQ Ultraviolet Lamp System