Input/Output Connections

The P300M/P300MQ/P300MT has three customer use interlock inputs. A fault condition on any of these inputs will cause the system to shut down; the fault must be corrected before the system can be restarted.

There must be one RF detector per system. The remaining interlock inputs are optional. The system blower and external interlock inputs in the P105A connector must be jumpered if these inputs are not used.

NOTE: All external switches must have gold contacts or be designed for use with logic-level signals to ensure reliable operation.

Table 5. J105A Inputs (Customer Use Interlocks)

Interlock

Pins

Fault

Contact

Application Comments

Condition

Ratings

 

 

 

 

 

 

 

 

RF Detector

1,2,3,4

RF trip or

12mA @12 VDC

Single lamp: One detector per system

(Master)

 

no detector

 

Multiple lamps: RF Master must have

 

 

present

 

RF detector.

 

 

 

 

 

RF Detector

3,4

OPEN

12mA @12 VDC

Single lamp: N/A

(Slave)

 

CIRCUIT

 

Multiple lamps: Typically jumpered. User

 

 

 

 

should remove jumper to install

 

 

 

 

additional RF detectors and set S1-5 to

 

 

 

 

ON (Yes)

 

 

 

 

 

SYS Blower

11,12

OPEN

12mA @12 VDC

Typically jumpered. User may remove

 

 

CIRCUIT

 

jumper to install N.O. pressure switch.

 

 

 

 

Multiple switches should be wired in

 

 

 

 

series.

 

 

 

 

 

External

9,10

OPEN

12mA @12 VDC

Typically jumpered. User may remove

Interlock

 

CIRCUIT

 

jumper to install N.O. interlock switch.

 

 

 

 

Multiple switches should be wired in

 

 

 

 

series.

 

 

 

 

 

Remote Inputs

One switch is required for remote operation of an F300S/F300SQ (to control the STANDBY/LAMP ON functions). Connections are made through the P106 connector. Single or Master/Slave systems may be operated remotely.

Installation, Operation, and Maintenance Manual # 509252 Rev R