Table 18. Troubleshooting/Fault Diagnosis (continued)

 

Signal/

 

 

Operation

Other

Possible Cause

Corrective Action

 

Symptoms

 

 

 

 

 

 

Switch from

LAMP OUT

Diodes, capacitors

If the readings are not different, a power

STANDBY to

indicator lights

 

supply fault is indicated. Disconnect all

ON

after 8

 

power and inspect for wiring damage. Test

(continued)

seconds;

 

diodes and capacitors according to the

Lamp bulb

 

procedures in Chapter 6, “Maintenance”.

 

 

 

does not light

 

 

 

Irradiator

Disconnect all power. Remove the

 

(continued)

 

 

irradiator cover (see Chapter 6,

 

 

 

 

 

 

“Maintenance”). Inspect for obvious

 

 

 

damage to wiring. Inspect the reflector for

 

 

 

pitting near the large slots, which is

 

 

 

evidence of arcing.

 

 

 

 

 

 

Magnetron

If no power supply fault can be found,

 

 

 

disconnect all power. Replace magnetron

 

 

 

(see Chapter 6, “Maintenance”).

 

 

 

 

WARNING: If no magnetron test voltage is measured at the magnetron test points, remove power from the unit immediately. Consult Fusion UV Systems Technical Services.

74F300S/F300SQ Ultraviolet Lamp System