Table 7. J105B Outputs (Customer Interface) (continued)

 

J105B

Asserted

Circuit Ratings

Comments

 

State

 

 

 

 

 

 

 

 

 

LAMP ON (K4)

9, 10

Closed

1 A @120 VAC

K4 contacts close when photocell

 

 

 

(10mA minimum)

is satisfied. Note: If using K4 for

 

 

 

 

press control, wait 6-8 seconds

 

 

 

 

for lamp output to stabilize.

 

 

 

 

 

PRESS ENABLE

11, 12

Closed

1 A @120 VAC

K5 contacts remain closed when

(K5)

 

 

(10mA minimum)

no unit or system error exists.

 

 

 

 

 

NOTE: When using circuit currents for K1-K5 below 10mA, contact Fusion UV Systems Customer Service.

Alternate Function for K5

An alternate K5 function may be selected by installing a jumper onto JP2 on the control card. The STANDBY-READY function is asserted (K5 closed) when the power supply completes the transition from the STANDBY-WAIT (flashing yellow) to the STANDBY-READY state (steady yellow). K5 remains closed for the LAMP ON state (flashing or steady green), the QRO state (alternate flashing of yellow and green), and the return to the STANDBY-READY state (steady yellow). K5 opens during the OFF/RESET state and the STANDBY-WAIT state (flashing yellow) from either the LAMP ON state or from the OFF/RESET state.

Table 8. Alternate Function for K5

 

J105B

Asserted State

Circuit Ratings

Comments

 

 

 

 

 

STANDBY

11, 12

Closed

1 A @120 VAC

K5 contacts close once

READY

 

 

(10mA minimum)

STANDBY-READY state is

(K5 alt.)

 

 

 

achieved, and for any higher

 

 

 

 

state.

 

 

 

 

 