Table 18. Troubleshooting/Fault Diagnosis (continued)

 

Signal/

 

 

Operation

Other

Possible Cause

Corrective Action

 

Symptoms

 

 

 

 

 

 

 

BLWR

Single-phase

See “Basic Checks” on page 65.

 

PRESSURE

voltage missing

 

 

indicator lights

phase

 

 

 

 

 

 

 

Air pressure

Measure air pressure inside irradiator

 

 

 

following the procedure described in

 

 

 

Chapter 6, “Maintenance”. IF PRESSURE

 

 

 

IS LOW:

 

 

 

 

 

 

Dirty air filters

Replace dirty air filter on remote or

 

 

 

modular blower.

 

 

 

 

 

 

Pressure switch

Replace the pressure switch.

 

 

faulty

 

 

 

 

 

 

 

Air handling design

Check for adequate air handling

Switch from

 

 

requirements.

 

 

 

STANDBY

HV cable

Connect HV cable.

OFF to

STANDBY

switch does

disconnected

 

 

not light

 

 

 

 

 

 

 

No fault

Faulty HV cable

Inspect cable, check for continuity.

 

indicators light.

 

 

 

Control card

Replace control card.

 

 

 

 

 

 

 

LAMP OUT

Card connectors not

Check control card. All connectors must be

 

and/or PWR

secure

properly secured.

 

IMBAL

 

 

 

Control card

Replace control card.

 

indicators light

 

 

 

 

immediately

 

 

 

 

 

 

 

RF INTLK

No RF detector

Connect RF detector. A system must have

 

indicator lights

connected

at least one RF detector connected.

 

or flashes

 

 

 

Programming error:

1. Close S1-5 on unit with RF detector

 

alternately with

 

1. All units have

connected.

 

the lamp

 

S1-5 open.

 

 

control

 

 

2. Unit(s) with S1-5

2. Install jumper.

 

switches

 

open and no jumper

 

 

 

 

 

 

between J105A-3

 

 

 

and J105A-4.

 

 

 

 

 

 

 

RF cable shorted or

Check cable for continuity.

 

 

open

 

 

 

 

 

 

 

RF detector faulty

Replace RF detector.

 

 

 

 

 

 

Control card

Replace control card.

 

 

 

 

Installation, Operation, and Maintenance Manual # 509252 Rev R