Table 18. Troubleshooting/Fault Diagnosis (continued)

 

Signal/

 

 

Operation

Other

Possible Cause

Corrective Action

 

Symptoms

 

 

 

 

 

 

Switch from

PS TEMP

Power supply fan

Repair or replace fan.

OFF to

indicator lights

inoperative

 

STANDBY or

 

 

 

 

Power supply too

Move to cooler location.

from STANDBY

 

 

hot or in poor

 

to ON

 

 

 

location

 

 

 

 

 

 

 

 

 

 

Temperature switch

With power off, check the T101 and T102

 

 

faulty or not

thermal switches, which should be closed,

 

 

connected.

with an ohmmeter or DVM. Check that the

 

 

 

thermal switch interconnections are

 

 

 

secure.

 

 

 

 

 

 

Card connectors not

Reseat connectors.

 

 

secure

 

 

 

 

 

 

 

Control card

Replace control card.

 

 

 

 

 

SYS BLWR

Blower power;

If SYS BLWR input for external blower in

 

indicator lights

pressure

J105A is in use, check external blower

 

 

 

power and pressure in the same manner

 

 

 

as remote blower.

 

 

 

 

 

 

Jumper in J105A

If SYS BLWR input not in use, check

 

 

 

jumper is in place in connector J105A.

 

 

 

 

 

 

Control card

Replace control card.

 

 

 

 

Lamp Control at

PWR ERROR

K101

Check for presence of magnetron current.

STANDBY

indicator lights

 

If within specifications, then examine

 

 

 

relays for stuck armature or contacts

 

 

 

welded shut. Check operation of relay with

 

 

 

single-phase power only connected.

 

 

 

 

 

 

Control card

Replace control card.

 

 

 

 

CAUTION: Disconnect single-phase power at main circuit breaker immediately. Magnetron may be severely damaged by application of high voltage when lamp control is in OFF or STANDBY.