HP E1433A User's Guide

Module Description

Status LEDs

￿Fail: This is the standard VXI “Failed” indicator. It lights briefly when powering up and normally goes out after a few seconds. If it stays on it indicates a hardware failure in the module.

￿Acs: This is the standard VXI “Access” indicator. When it is on, it indicates that another device on the bus is contacting the module, for example to transfer data or read registers.

￿Trigger: This LED flashes on each time the measurement triggers, so when it is blinking it indicates that the measurement is triggering.

If your HP E1433A has the Tachometer option, this LED is defined differently. See the chapter: The Tachometer Option (AYF).

SMB Connectors

￿ExSamp: This is an input connector for an external sample clock. The sample clock must be TTL level and have a frequency between 40.96 kHz and 100 kHz. Internally this frequency can be decimated.

￿Cal: This connector is used for calibration. It can be configured to output a calibration signal or to accept an input calibration signal. See the calibration section in this chapter.

￿ExTrig: This allows for an external trigger input to the HP E1433A. The input signal must be TTL, other characteristics can be defined in software. ExTrig can be enabled or disabled in software.

Input Connectors (1 or 2)

These connectors are attached to the cables from an 8-Channel Input (breakout box). They connect the input signal to the HP E1433A. Each connector carries four channels. Depending on options, there can be 1 or 2 input connectors (4 - 8 channels).

5-7

Page 113
Image 113
HP 8-Channel 196 kSa/sec Digitizer plus DSP E1433A manual Status LEDs