Using the HP E1433A

Managing Multiple-mainframe Measurements

In a single-mainframe measurement, the HP E1433A communicates with other HP E1433As through the TTLTRG lines. However, when using the VXI-MXI bus extender modules, the TTLTRG lines, which carry the group synchronization pulse and sample clock, are extended only in one direction. This unidirectional signal connection restricts the types of measurements you can make in a multiple mainframe environment.

You cannot perform the following types of multiple mainframe measurements:

￿Unequal pre-trigger delay settings between mainframes

￿Channel triggering by channels in Mainframe B

￿Lower spans or longer blocksizes in Mainframe B

￿Different digital filter settling times between HP E1433A modules

Slot 0

HP E1482B

 

 

HP E1433A

Contoller

 

 

 

 

 

 

Fail Acs Trigger Fail Acs

Trigger

 

 

ExSamp Cal ExTrig ExSamp Cal ExTrig

 

 

8-CHANNEL 196

8-CHANNEL 196

 

 

kSa/s DIGITIZER

kSa/s DIGITIZER

 

 

+DSP

+DSP

 

 

 

Chan

Chan

 

 

 

5-8

5-8

 

 

 

Chan

 

Chan

 

 

1-4

1-4

 

 

HP E1433A

HP E1433A

VXI Mainframe A

Fail Acs Trigger Fail Acs Trigger

 

ExSamp Cal ExTrig ExSamp Cal ExTrig

 

8-CHANNEL 196

8-CHANNEL 196

 

kSa/s DIGITIZER

kSa/s DIGITIZER

 

+DSP

+DSP

 

Chan

Chan

 

5-8

5-8

 

Chan

Chan

 

1-4

1-4

 

HP E1433A

HP E1433A

VXI Mainframe B

 

 

HP E1482B

 

HP E1433A

 

 

Figure 3-3: Multiple mainframes - two mainframes

3-11

Page 49
Image 49
HP 8-Channel 196 kSa/sec Digitizer plus DSP E1433A manual Managing Multiple-mainframe Measurements