HP E1433A User's Guide
Break Out Boxes
Recommendations on wiring for the E1432/3A 4 Channel Input Connector
Allowed Connections
Differential Input Channels
Connect at E1432/3A end of cabling and at DUT
Recommended: shielded twisted pair
1+ Diff 4
2- Diff 4
5+ Diff 3
6- Diff 3
8+ Diff 2
9- Diff 2
12+ Diff 1
13- Diff 1
Input Channel Shielding
Connect at E1432/3A end of cabling ONLY
3RFI GND/Drain Shield 4
4RFI GND/Drain Shield 3
10RFI GND/Drain shield 2
11RFI GND/Drain Shield 1
Additional shielding of entire cable
GND for grounded measurements if required
14 | RFI GND/Cable Shield |
26 | RFI GND/Cable Shield |