HP E1433A User's Guide
Using the HP E1433A
In the example above, Mainframe A contains the Slot 0 Controller for a multiple mainframe system. Mainframe A is connected to Mainframe B with a
Locate modules with logical addresses less than 128 in Mainframe A.
Locate modules with logical addresses greater than 127 in Mainframe B.
Locate the
Locate the last module in the module list specified in the call to hpe1432_init in Mainframe A.
Locate the module that generates the group synchronization pulse in Mainframe A.
Locate the channels performing channel triggering in Mainframe A.
Locate the module with the shared sample clock in Mainframe A.
If you do not use a groupID with the call hpe1432_readRawData or
hpe1432_readFloat64Data, empty the HP E1433As’ FIFOs in Mainframe B before Mainframe A. In other words, do not empty the FIFOs in Mainframe A unless you have emptied the FIFOs in Mainframe B. For more information about groupID see “Grouping of Channels/Modules” in this chapter.
If more than two mainframes are needed,
Phase Performance in Multiple Mainframe Measurements
Phase specifications are degraded by the delay that the
Using a
The cable adds approximately 6 ns per meter of cable.
Each