HP VXI E1432A manual Providing an External Clock

Models: VXI E1432A

1 220
Download 220 pages 26.29 Kb
Page 118
Image 118

HP E1432A User's Guide

Module Description

Providing an External Clock

You can program the HP E1432A to accept an external word rate clock from the Sample 0 line on the VXI Bus. The digital filters are still functional, providing a range of effective word rates. All sampling is done simultaneously and is not multiplexed.

To connect an External Sample Clock, use the External Sample SMB connector on front panel of the HP E1432A. External Sample at word rate and External Trigger are available on the front panel of HP E1432A’s which do not have an arbitrary source or tachometer option.

The external clock must be a fixed frequency. Its maximum frequency must not be higher than 100 kHz. Its minimum frequency must be at least 40.96 kHz.

5-14

Page 118
Image 118
HP VXI E1432A manual Providing an External Clock