Using the HP E1432A

Arm and Trigger

This section explains some terminology relating the the “Arm” and “Trigger” steps in the measurement loop. As an example a measurement might be set up to arm at a certain RPM level and then subsequently trigger at an external event corresponding to top dead center (TDC). The settings would be:

￿

Arm:

RPM Step Arm

￿

Trigger:

External Trigger

If you want to begin a throughput session at this same RPM/TDC event, then the first external trigger after a specified RPM would start a continuous mode measurement. Now (using overlap block mode) the settings would be:

￿ Pre-Arm:

RPM Step Arm

￿ Arm:

Auto

￿ Trigger:

Auto

In the measurement loop, an arm must take place before a trigger. You can program how many triggers to do before waiting for another arm condition. The default is one trigger for each arm. For each trigger, a block of data is sent to the host.

The first arm in a measurement is the pre-arm. By default, the pre-arm condition is the same as the regular arm conditions.

Valid Arm (and Pre-Arm) conditions are:

￿Auto Arm

￿Manual Arm

￿RPM Step Arm

Valid trigger conditions are:

￿Auto Trigger

￿Input Trigger

￿Source Trigger

￿External Trigger

￿Manual Trigger

￿Tachometer Edge Trigger

3-24