Service Manual M695

Testset

Doc code 638 718

09/93 3077-HWE

 

 

 

Contents of test set

Enables defective assemblies to be identified

Procedure for fault location

See section 4

Cause in footswitch?

Cause in microscope of in XY-gear?

Cause in control unit or swingarm?

Test set M695/M680 - stock no. 566 152

Test box, stockno. 566 146

Service cable M695/M680, stock no. 566 148

Service handswitch, stock no. 566 147

Foot- or handswitch

S w i n g a n n

Control unit MEL61/62 and MEL64/65

X- or Y-gear MSV138 (M680 only)

M695/M680 microscope

1.Introduce the test box between the swingarm and the microscope - Plug connections JP36 and JP37.

2.Inspect the voltages and currents of the POWER SUPPLY as described in Section 4.

3.Substitute service handswitch for footswitch, observe signal voltages, and also monitor LEDs on control unit.

4.Observe the signal voltages of the defective function (MICROSCOPE CONTROL or XY-gear) - as described in Section 4.

5.Separate the microscope or the XY-gear, as appropriate, from the test box, and observe the signal voltages which flow from the control unit through the swingann.

6.Use service cable to bridge swingarm, and observe signal voltages from control unit.