Description of control unit MEL64/65 | Service manual M695 |
03/93 | Doc code 636 718 |
|
|
ControI LEDs | The function display | tells you whether the | signals from the footswitch or |
| handswitch are being | correctly received and | interpreted: |
for FOCUS UP
for FOCUS DOWN
for CENTER POSITION
Motorized focusing | The handswitch or footswitch transmits on RD_2 or RD_3 the codes which |
| correspond to the functions: |
| Data format: |
ASCII
repeating at
The logics on the receiver board (processor) assign the received codes to their appropriate functions:
FOCUS UP
FOCUS DOWN
CENTER POSITION
The control signals (0V/5V) for FOCUS UP and FOCUS DOWN reach the logics board in the microscope via the swingarm cable.
The voltage level of FOCUS SPEED depends on the setting of the potentiometer on the panel board.
|
|
| min. FOCUS SPEED | max. FOCUS SPEED |
|
|
|
|
|
Zoom | position | 25 | 2.1V | |
|
|
|
|
|
Zoom | position | <8 | 5V | |
|
|
|
|
|