Service manual M695

Modification

Doc Code 636 718

10/94 3077-HWE

Changes and modifications

M680 control unit (MEL61/62) is compatible withM695 surgical microscope

. . . from control unit the MEL61/62 can also be used for the M695, provided that earlier lamp

serial number 011193001 control PCBs are subjected to a slight modification. This modification was

onwards introduced from March 1993 onwards (Index C on board) (355 980 C).

The functions ZOOM SPEED B, XY-SPEED and POSITION 1/2 are not required for the M695. The positions occupied on the control panel by these switches or potentiometers will be covered over. The covers, and instructions for fitting them, accompany the control unit.

From March 1995 onwards, only the MEL61/62 control unit will be used for the M680 and M695. The previously-used MEL64/65 control unit will no longer be manufactured.

Modification of the is required if both of the following conditions are met: lamp-conbol PCB

1.The MEL64/65 control unit is replaced by an MEL61/62

2.The M695 was despatched before March 1993. The lamp control PCB therefore bears an earlier index than C.

Under these conditions the hand-/footswitches will not react at the control input A or B respectively, depending on the position of the switch "Position 1/2".

Modification: On the soldered side, bridge pins 10 and 12 of P20 with a 100 Ω resistor (1/8W). The resistor shall not come into contact with the housing cover.

2-1