Inspection procedure | Service Manual M695 |
09/93 | Doc Code 636 718 |
2.1Turn LAMP potentiometer on control unit from min. to max.
• | Reading on test box: LAMP REF. | min./max. 0.3......5v. |
| The brightness of the microscope lamp is continuously adjustable. |
2.2Activate the
•The spare bulb is moved into the beam path, and lights up.
•The orange indicator springs out.
(Push the orange indicator back into the rear wall)
(Fault location in lamp control: page
3.1Turn the ZOOM SPEED potentiometer from min. to max. and from max. to min.
• | Reading on test box: ZOOM SPEED | min./max 1.84...2.34Vl5V ±5% |
| The zoom speed is continuously adjustable. |
|
3.2Use
• | Reading on test box: ZOOM UP | Low/High <0.2V/>4.75V |
3.3Use
• | Reading on test box: ZOOM DN | Low/High <0.2V/>4.75V |
(Fault location in ZOOM control: page
3.4Mechanical zoom movement.
• Zoom moves uniformly in both directions.