Service manual M695

M695 Fault finding

Doc code 636 718

09/93 3077-HWE

 

 

 

Focus control

Select microscope side A or B:

Select A or B, depending on which side the zoom works incorrect

Select A if the fault appeares on both sides, A and B.

 

Action

Reading

Reaction

Possible fault

 

 

 

 

 

 

 

FOCUS SPEED set min. focus speed

 

 

 

1.25 . . . 1.27V

 

 

 

 

 

set max. focus speed

5v +-5%

 

 

 

 

FOCUS UP

 

 

 

 

 

 

/footswitch

>4.75V

 

 

Voltage ok, but

Microscope

 

 

<0.2V

 

 

 

non-actuated

 

 

reacts incorrect

(Focus servo,

 

 

 

 

 

 

wiring)

FOCUS DN

 

 

 

 

 

 

/footswitch

>4.75V

 

 

 

 

 

non-actuated

<0.2V

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Separate microscope

see above

 

 

 

from the testbox and

Voltages ok

Microscope

 

check signals

 

 

 

 

(short circuit on

 

 

 

 

 

 

Focus servo or

 

 

 

 

 

 

wiring)

 

 

 

 

 

 

 

 

Bridge swingarm with

 

 

 

 

 

 

servicecable and verify

see above

Voltages ok

Swingarm-

 

signals

 

 

 

 

cable

Voltages not ok

Control unit