Service manual M695 | M695 Fault finding | |
Doc code 636 718 | 09/93 | |
|
|
|
Focus control
Select microscope side A or B:•Select A or B, depending on which side the zoom works incorrect
•Select A if the fault appeares on both sides, A and B.
| Action | Reading | Reaction | Possible fault | ||
|
|
|
|
|
|
|
FOCUS SPEED set min. focus speed |
|
|
| |||
1.25 . . . 1.27V |
|
|
|
| ||
| set max. focus speed | 5v |
|
|
|
|
FOCUS UP |
|
|
|
|
| |
| /footswitch | >4.75V |
|
| Voltage ok, but | Microscope |
|
| <0.2V |
|
| ||
|
|
| reacts incorrect | (Focus servo, | ||
|
|
|
|
|
| wiring) |
FOCUS DN |
|
|
|
|
| |
| /footswitch | >4.75V |
|
|
|
|
| <0.2V |
|
|
|
| |
|
|
|
| |||
|
|
|
|
|
|
|
| Separate microscope | see above |
|
| ||
| from the testbox and | Voltages ok | Microscope | |||
| check signals |
|
|
|
| (short circuit on |
|
|
|
|
|
| Focus servo or |
|
|
|
|
|
| wiring) |
|
|
|
|
|
|
|
| Bridge swingarm with |
|
|
|
|
|
| servicecable and verify | see above | Voltages ok | Swingarm- | ||
| signals |
|
|
|
| cable |
Voltages not ok | Control unit |