M695 Fault finding

Service manual M695

09/93 3077-HWE

Doc code 636 718

 

 

Zoom control

 

Action

 

 

Heading

Reaction

 

Possible fault

 

 

 

 

 

 

ZOOM SPEED Set min. zoom speed

1.84 ... 2.34V

 

 

 

 

Set max. zoom speed

5V +-5%

 

 

 

ZOOM UP

 

 

>4.75V

 

 

 

 

/footswitch

 

Voltages ok, but

Microscope

 

 

 

 

 

 

non-actuated

 

<0.2V

reacts incorrect

(Logic board.

 

 

 

 

 

 

 

Zoom servo,

 

 

 

 

 

 

 

wiring)

ZOOM DN

 

 

>4.75v

 

 

 

 

/footswitch

 

 

 

 

 

non-actuated

 

<0.2v

 

 

 

 

Separate microscope

see above

Voltages

ok

Microscope

 

from the testbox and

 

check

signals

 

 

 

 

(short circuit on

 

 

 

 

 

 

 

Zoom servo or

 

 

 

 

 

 

 

wiring)

 

Bridge

swingarm

with

 

 

 

 

 

servicecable and

verify

see above

Voltages

ok

Swingarm cable

 

signals

 

 

 

 

 

 

 

 

 

 

 

Voltages not ok

Control unit

 

 

 

 

 

 

 

 

4-8