Service manual M695

Description swingarm MSV136/137

Doc code 636 718

09/93 3077-HWE

 

 

List of signals for swingarm MSV 136/137

Required values for standard two-metre swingarm cable

Unless otherwise indicated, set the lamp to minimum brightness, because the values quoted are only valid under these conditions.

Power supply
SIGNALVALUE

TOL.

Voltage against PGND

26V

+-5%

Lamp min. brightness ->

Voltage against PGND

20V

+-5%

Lamp max. brightness ->

Voltage against SENSE

-50mV

 

Lamp min. brightness ->

 

Voltage against SENSE

-250mV

 

Lamp max. brightness ->

 

Voltage drop against logics level

4.9V

-0.2v

Lamp min. brightness: ->

(Test box: Microscope

 

 

Setting: Focus up)

 

 

Voltage drop against logics level

4.2V

-0.8V

Lamp max. brightness ->

(Test box: Microscope

 

 

Setting: Focus up)

 

 

6-1