Service manual M695

Description of control unit MEL64/65

Doc code 636 718

09/93 3077-HWE

 

 

 

Travelling to mid-positionThis function is activated from the handswitches or footswitches A and B, using the key (SL->). A timer (Monoflop) sets FOCUS SPEED to GND potential for about 30 milliseconds, using a transistor.

The impulse is recognized on the focus servo in the motor driven objectiv (MOBxxx), and the system is steered to the mid-position.