Service manual M695

M695 Fault finding

Doc code 636 718

09/93 3077-HWE

 

 

 

Power supply of control unit

Action

 

 

ReactionPossible reasonCheck

 

 

 

 

 

 

SWITCHING ON

 

 

Lamp of main switch

Line voltage

Fuses F1 and F2 in

 

 

 

does not light

 

line plug P1

 

 

 

 

 

Main switch S1

 

 

 

 

 

Line cable

 

 

 

 

 

 

 

 

 

5V-LED on Function-

24V-main supply

Transformer

 

 

 

Display does not light

 

Rectifier circuit

 

 

 

 

 

Wiring

 

 

 

 

10V-supply

Panel board

 

 

 

 

5V-supply

Receiver board

 

 

 

 

 

 

 

 

 

Focus does not move

Control unit,

Focus control

 

 

 

to center position

Swingaxm o r

page 4-9

 

 

 

 

microscope

 

 

 

 

 

 

 

4-3