Service manual M695 | M695 Fault finding | |
Doc code 636 718 | 09/93 | |
|
|
|
M695 Fault finding
Preparation
1. Connecting the testbox
•
•
•
•
Disconnect plug P36 from swingarm
Hook in the testbox to the swingarm and plug in P36 at the testbox.
Connect the testbox cable (green sleeve) with J36 at the swingann. With that the control unit is connected with the microscope through the testbox.
Set switch “MICROSCOPE SELECT” on the testbox to position A.