Inspection procedure

Service Manual M695

09/93 3691-DMF

Doc Code 636 718

1.3Activate the foot- or handswitch functions.

The appropriate LED on the function display shines.

(If a fault arises, exchange the foot- or handswitch and repeat the inspection procedure. If there is still a fault, then its source is to be found in the MEL64/65 control unit.)

(Fault location in hand-/footswitch functions, page 4-4)

ML on/off

M i c r o s c o p e l a m p M L

A on/off

Additional illumination A

Focus up

Focus down

Zoom up

Zoom down

1.4Connect cables of microscope and swingarm,

• The plugs must go in and out without difficulty.