Inspection procedure | Service Manual M695 |
09/93 | Doc Code 636 718 |
1.3Activate the foot- or handswitch functions.
•The appropriate LED on the function display shines.
(If a fault arises, exchange the foot- or handswitch and repeat the inspection procedure. If there is still a fault, then its source is to be found in the MEL64/65 control unit.)
(Fault location in
ML on/off
M i c r o s c o p e l a m p M L
A on/off
Additional illumination A
Focus up
Focus down
Zoom up
Zoom down
1.4Connect cables of microscope and swingarm,
• The plugs must go in and out without difficulty.