M695 Fault finding

Service manual M695

09/93 3077-HWE

Doc code 636 718

 

 

Check lines of Powerground PGND

(Voltage drop on PGND relating to logic ground)

Signal

Action

 

Reading

Reaction

Possible fault

 

 

 

 

 

FOCUS UP

Adjust lamp to maximal

max. -0.8W

If voltage is ok - depending on the incorrect

 

brightness

with standard-

function, check signals of:

 

 

swingarm

 

Lamp control - page 4-7

 

 

cable

 

Zoom control - page 4-8

 

 

 

 

 

Focus control - page 4-9

 

 

 

 

 

 

 

Bridge swingarm with

 

max. -0.8V

Voltage ok

Swingarm cable

 

service cable and adjust

 

 

 

(One of the two

 

lamp to max. brightness

 

 

 

PGND-lines is

 

 

 

 

 

broken)

 

 

 

 

Voltage not ok

One of the two

 

 

 

 

typ. >-1v

PGND-lines in

 

 

 

 

 

the microscope

or control unit is broken.

Important: Adjust lamp again to minimal brightness!