SOM and SM Service News

Surgical Operating Microscopes

No. 10/95

To Service staff

Date/from June 28, 1995 / 3077-HWE

File Workshop binder “Service News” / Service manual M695

Re. M695 surgical microscope in conjunction with multifocus objective 445 944

Leica AG

CH-9435 Heerbrugg

(Switzerland)

Restriction of illuminated area

lnspect the illumination

Multifocus objective on M695:Readjustment of the illumination

When the multifocus objective is used with the M695, the illuminated area available to the surgeon may be shadowed or vignetted even when the the lamp is correctly adjusted.

When you use the multifocus objective for the first time, inspect the illumination:

Engage the shortest working distance (the shortest focal length).

Engage the lowest zoom position (7).

Inspect the illuminated fields produced by both of the lamps in the lamp changer.

If the illuminated area is noticeably restricted, readjust the lamps.

Readjusting the lamp

The illumination of the field of view is influenced by the distance

 

between the lamp and the condenser lens. This distance can be changed

 

by means of the eccentric stop cam on the lamp changer. To obtain

 

uniform illumination, reduce the distance between the lamp and the

 

condenser lens.

 

Adjust the eccentric stop cam for both lamps in the lamp changer:

 

1. Remove the multifocus objective and push the filter slide of the

 

M695 optics carrier towards the objective. Beneath the filter slide

 

you will find the adjusting screw for the eccentric stop cam.