Service manual M695 | M695 Fault finding |
Doc code 636 718 | 09/93 |
|
|
Voltage | Action | R e a d i n g | Reaction | Possible fault |
|
|
|
|
|
24V M I C R O | 1.Lamp to min.brightness |
| 2.Lamp to max.brightness |
| 3.Lamp back to min. |
| brightness |
26V | Microscope reacts - | Check supply- | |
20V | but incorrect | lines of 24V and | |
|
|
| check PGND - |
|
|
| page |
|
| Microscope without | Wiring of |
|
| any function | microscope |
Switch off, bridge swing- | microscope reacts | Swingarm cable |
arm with servicecable and 26V | ||
switch on again |
|
|
|
|
|
Switch off, | Voltage ok | evtl. short circuit |
Separate microscope from 26V | ||
testbox and switch on |
| in the microscope |
| Voltage not ok | Control unit |