Testset | Service | Manual M695 | |
09/93 | Doc | code 636 718 | |
|
|
|
|
Functions of test box
All control voltages . . . | All supply- and control voltages between the control unit and the microscope |
| with |
. . . to the microscope . . . | Control voltages to the microscope are measured relative to XY_GND, such |
| to the |
. . . measure at low light | Unless otherwise indicated, set the lamp to minimum brightness. The voltages |
level | indicated in Section 4 are only valid for this condition. |
| Voltage levels are falsfied by the voltage drop between the control unit and |
| the microscope. Because of the current taken by the lamp, this voltage drop |
| to PGND (measuring ground) can be as much as 0.9V. The control voltages |
| of focus and zoom are measured against PGND and are therefore influenced. |