Description of control unit MEL64/65 | Service | manual M695 | |
09/93 | Doc | code 636 718 | |
|
|
|
|
The function display tells you whether the signals from the footswitch or handswitch are being correctly received and interpreted:
for ZOOM UP
for ZOOM DOWN
The handswitch or the footswitch transmits on RD_2 or RD_3 the codes which correspond to the functions:
Data format:
ASCII
repeating at
The logics on the receiver board (processor) assign the codes received to their appropriate functions:
ZOOM UP
ZOOM DOWN
The control signals (0V/5V) for ZOOM UP and ZOOM DOWN reach the logics board in the microscope via the swingarm cable.
The voltage level of ZOOM SPEED depends on the position of the potentiometer on the panel board.
| Level of ZOOM SPEED |
min. ZOOM SPEED | 2.10V +- 0.25V |
|
|
max. ZOOM SPEED | 5V +- 5% |
|
|