Drainage

Plumbing

Drainage

The drainage system is fitted with a non-return valve which prevents dirty water from flowing back into the machine.

The machine should preferably be connected to a separate drainage system on-site.

If no separate drainage system is available, contact the Miele Service Department.

The on-site drain connection point should be located between 0.3 m and 1 m above the lower edge of the machine.

If the connection is lower than 0.3 m, the hose should loop at a height

of at least 0.3 m to ensure correct drainage.

The drainage system must have the capacity to take a minimum drainage flow of 16 litres/minute.

The machine is supplied with approx. 1.4 m of flexible drain hose with an internal diameter of 22 mm. Two hose clips are also supplied for connection.

The drain hose must not be shortened.

A longer drain hose (up to 4 m long) is available to order from the Miele Spare Parts Dept.

The drainage system must not ex- ceed 4 metres.

See also the installation diagram supplied.