IV Connecting and Installing the Equipment

2 Installing the DS-L1

2Installing the DS-L1

When installing the DS-L1 on a flat surface such as a desktop by using the stand arm

Extract the stand arm from the DS-L1 and place it at an angle of 5 to 10 degrees, as shown below.

Leave a clearance of about 100 mm around the DS-L1 to prevent heat from collecting near the equipment.

When attaching a commercially available arm to the DS-L1

Any commercially available VESA standard-compliant 75-mm arm, UL1678-suited stand and mount can be attached to the DS-L1. When doing so, make sure the following specifications are met:

DS-L1 weight:

1.3 kg

Fitting screw hole:

M4, 7-mm deep

Tightening torque:

80-120 N·cm

Leave a clearance of about 100 mm around the DS-L1 to prevent heat from collecting near the equipment.

M4 screw holes

Arm

Side view of the DS-L1

Rear view of the DS-L1