Site Power Wiring Recommendations

The table below describes the conditions{ XE "Scanner installation:Site power wiring recommendations" } { XE "Power wiring recommendations" }for power wiring at a scanner site.

Condition

Description

Dedicated Feeder

Run a separate dedicated feeder from the building main power panel to

 

the branch power panel serving the 5000i system. The cabling should be

 

of sufficient diameter to produce no less than 110V at the outlet when

 

the 5000i is connected and powered on. This feeder should be a

 

minimum of 30 amp service to allow for future growth of the system.

 

 

Dedicated Isolated/

An isolated/dedicated ground is required. Run this ground conductor as

Insulated Ground

an uninterrupted circuit from the power panel to the 5000i system branch

 

panel. The gauge of this conductor must be at least as large as that of

 

the AC circuit conductors in all stages of the circuit.

 

 

Panel Branch Circuits

All branch circuits leaving tne 5000i branch power panel should be at

 

least 10 AWG copper wire. There should be no circuits from this power

 

panel serving devices that are not part of the 5000i system.

 

One branch circuit from this power panel should be dedicated at the

 

scanner (this circuit will also service the computer and monitor), and

 

there should be at least one additional dedicated circuit for each

 

additional scanner.

 

 

Additional Utility

Provide additional convenience outlets in the vicinity of the 5000i

Convenience Connectors

scanner. Connect these outlets (may be used for vacuums, paper

 

joggers, and other electrical devices) to a branch panel other than the

 

scanner branch panel to provide maximum noise isolation for the 5000i

 

system components.

 

 

Scanner Electrical

Dedicate one 15 amp (120 VAC) or 7.5 amp (220 VAC) circuit to each

Connections

5000i scanner at the site. All may be served from the same branch

 

panel, provided its rated capacity is not exceeded.

 

 

Electrical Outlets{ XE

All 120V outlets for the 5000i scanner should be Hubble IG-5362 or

"Electrical outlet" \i }

equivalent (NEMA 5-20R). All 240V outlets should be Hubble IG2410A

 

or equivalent. Outlets for the peripheral equipment should be of the type

 

and rating specified in the hardware reference manuals for the

 

equipment.

 

The quantity of outlets that must be provided for the peripheral

 

equipment varies according to system configuration. (See Panel Branch

 

circuits above.) Outlets should be located within 6 feet (2 meters) of the

 

working location of the equipment.

 

 