2.3.1Adjusting Emissivity

2Using the Handheld Infrared Thermometer

2.3.1Adjusting Emissivity

°F

Refer to Appendices B and C for information on emissivity.

1.Determine the emissivity of the target.

2.Aim at the target and pull the trigger.

3.If necessary, press the Refer to Appendices B and C for information on emissivity. key to increment the target

emissivity or press the 2.3.2 Using the LOCK Function key to decrement the target emissivity.

NOTE

The Emissivity Display Mode (E) appears every time the trigger is pulled regardless of how the Display Mode was previously set.

NOTE

The emissivity setting does not change when the thermometer is turned off. However, when the batteries are replaced, the emissivity is reset to 0.95, the default value.

2.3.2 Using the LOCK Function

This function electronically locks the trigger mechanism:

1.Pull the trigger and press the This function electronically locks the trigger mechanism: key to lock the trigger. The

Manual background icon will appear on the display.

2.Release the trigger. This allows the thermometer to operate continuously whether or not the trigger is pulled.

NOTE

To unlock the trigger function, press the Manual background key again,

and the Manual background icon is no longer displayed.