Omega Engineering OS530HR, OS531, OS53X-CF, OS524, OS533 Troubleshooting Guide, Problem, Solution

Models: OS524

1 78
Download 78 pages 980 b
Page 54
Image 54
Troubleshooting Guide

 

 

Troubleshooting Guide

6

 

 

 

 

 

 

 

THERMOMETER

 

 

 

 

 

 

 

 

 

 

 

Problem

Solution

 

 

 

 

The thermometer does 1a.

Properly install fresh batteries.

 

 

 

 

not turn on (No Display)

 

 

 

 

 

1b.

If operating under ac power, check

 

 

 

that the ac adapter is plugged in

 

 

 

properly to the ac wall outlet and to the

 

 

 

thermometer.

 

 

 

 

1c.

Make sure the batteries make good

 

 

 

contact - remove and reinstall the

 

 

 

batteries.

 

 

2.

Make sure that the trigger is pulled

 

 

 

completely.

 

 

 

 

 

 

 

 

1.Reset the thermometer. It sets all of the parameters to the default values and restores calibration. The procedure is as follows, when the thermometer is in sleep mode:

a.Press and hold the Problem key.

b.Pull the trigger.

c.Release the trigger.

d.Release the Solution key.

e.The version of the software is displayed for about 1 second. The emissivity display mode immediately follows with the emissivity of 0.95.

- The

 

icon

1.

Properly install fresh batteries.

flashes.

-The thermometer beeps intermittently.

-The thermometer

flashes Manual backgroundManual backgroundManual backgroundManual backgroundManual backgroundManual backgroundin the Main Display.

6-1

Page 54
Image 54
Omega Engineering OS530HR, OS531, OS53X-CF, OS524, OS533, OS532, OS530L, OS523, OS534 Troubleshooting Guide, Problem, Solution