5.1.3.1 PC-206 Amplifier/Multiplexer Board (9016)

The following procedures should be used for replacement of the PC-206 Amplifier/Multiplexer Board. Use the tools and follow the general precautions described in Section 5.1.1.

(1)Disassemble the module as described in Section 5.1.2. Carefully remove the wiring harness from connector P1 of the PC-206 board. Note the orientation of the PC-206 relative to the rest of the module to ensure the new PC-206 is installed in the same position.

(2)Remove the two (2) Phillips-head screws securing the PC-206 board to the DH200 transducers. Carefully disconnect the PC-206 board from the DH200s by slowly working the board off; starting at one end and moving down the length of the board. It is important that the gold pins are not bent when removing the board.

(3)Replace the old PC-206 board with a new one by placing it loosely on top of the DH200s. Ensure the board end containing connector P1 is oriented the same as the board just removed. Inspect and make sure that all the gold pins fit easily into the female end of the connector on the DH200 transducers. Press the board down evenly until all pins are firmly seated.

(4)Install the two (2) Phillips-head screws to secure the PC-206 to the DH200s. Be careful not to over-tighten. Install the wiring harness to connector P1 of the PC-206, ensuring proper pin 1 location. (Pin 1 of the ribbon cable has a red stripe while pin one of P1 will contain a square solder pad on the PC-206.)

(5)Carefully align the gasket on the top plate, ensuring it is free of contaminants. Re-install the module electronics into the extrusion case. Ensure that the alignment posts in the module's bottom panel align with the PC-203 or PC-280 electronics support brackets when placing the top panel and electronics back into the housing.

(6)Replace the screws that secure the top panel to the scanner housing and tighten. Do not over- tighten; 7-9 inch-pounds torque should be sufficient.

(7)Test your scanner to ensure proper operation.

5 - 6www.PressureSystems.com

Page 140
Image 140
Pressure Systems 9022 user manual 3.1 PC-206 Amplifier/Multiplexer Board

9022 specifications

Pressure Systems 9022 is a state-of-the-art pressure management system designed for various industrial applications, offering unparalleled efficiency and reliability. This innovative system has become a benchmark in pressure technology, providing users with advanced control and monitoring features.

One of the main features of Pressure Systems 9022 is its ability to deliver precise pressure regulation. Utilizing advanced algorithms, the system continuously monitors pressure levels and adjusts them in real-time, ensuring optimal performance and safety. This feature is particularly beneficial in industries such as oil and gas, pharmaceuticals, and food processing, where maintaining specific pressure levels is crucial for product integrity and process efficiency.

The 9022 is also equipped with cutting-edge connectivity options. With built-in IoT capabilities, the system can be easily integrated into existing networks, allowing for remote monitoring and control. Users can access pressure data in real-time through a user-friendly interface, which enhances decision-making and reduces response times in case of irregular conditions.

Another standout characteristic is its robust construction. Designed to endure harsh environmental conditions, the Pressure Systems 9022 is built from high-grade materials that provide excellent resistance to wear and corrosion. This attribute ensures longevity and minimizes maintenance costs, making it an economical choice for businesses.

Moreover, the system incorporates various safety features. Automatic shut-off valves and alarms are integrated into the design to prevent overpressure situations, protecting both equipment and personnel. These built-in safeguards align with industry standards and promote a culture of safety within organizations.

The Pressure Systems 9022 also emphasizes energy efficiency. By optimizing the pressure regulation process, it reduces energy consumption, contributing to lower operational costs and a smaller environmental footprint. This focus on sustainability resonates with modern businesses looking to enhance their eco-friendly practices.

In conclusion, Pressure Systems 9022 stands out as a versatile, efficient, and reliable solution for pressure management. Its advanced features, durability, and commitment to safety make it a valuable asset for any industrial operation. By adopting this technology, organizations can enhance performance while ensuring compliance with safety regulations, ultimately leading to a more productive and sustainable operation.