Pressure Systems, Inc.

NetScanner™ System (9016, 9021, & 9022) User’s Manual

5.2 9022 Excitation Trim

The 9022 output voltage is factory set at 10 volts. The following procedure can be used to trim the 10 volt output (~±5%).

(1)Disassemble the unit as described in Section 5.1.2.

(2)Locate R-19, the output voltage trim potentiometer, on the PC-317 board. It is the only potentiometer on the board and is near the end of the unit on the top side of the lower PC board (PC-317), toward the end of the unit nearest the main power and communications connector. Note that R-19 is a four turn potentiometer. See Figure 5.8, below.

Figure 5.8

PC-317 Board (Trim Potentiometer and Jumper

(3)Place the disassembled unit on a nonconductive surface. Using a spare connector, connect a calibrated voltage meter to pins G and H (+ and - respectively) of any of the transducer connectors.

(4)Apply power to the unit through the main power and communications connector. The 10 volt output voltage may be trimmed by rotating the potentiometer R-19.

(5)Once the output voltage is set, remove power from the unit.

(6)Reinsert the electronics into the housing, ensuring that the alignment posts in the module’s bottom panel align with the holes in the PC-280 mounting brackets. This may require loosening the three (3) screws attaching the PC-280 board to the top plate

5 - 19www.PressureSystems.com

Page 153
Image 153
Pressure Systems 9022 user manual Excitation Trim, PC-317 Board Trim Potentiometer and Jumper

9022 specifications

Pressure Systems 9022 is a state-of-the-art pressure management system designed for various industrial applications, offering unparalleled efficiency and reliability. This innovative system has become a benchmark in pressure technology, providing users with advanced control and monitoring features.

One of the main features of Pressure Systems 9022 is its ability to deliver precise pressure regulation. Utilizing advanced algorithms, the system continuously monitors pressure levels and adjusts them in real-time, ensuring optimal performance and safety. This feature is particularly beneficial in industries such as oil and gas, pharmaceuticals, and food processing, where maintaining specific pressure levels is crucial for product integrity and process efficiency.

The 9022 is also equipped with cutting-edge connectivity options. With built-in IoT capabilities, the system can be easily integrated into existing networks, allowing for remote monitoring and control. Users can access pressure data in real-time through a user-friendly interface, which enhances decision-making and reduces response times in case of irregular conditions.

Another standout characteristic is its robust construction. Designed to endure harsh environmental conditions, the Pressure Systems 9022 is built from high-grade materials that provide excellent resistance to wear and corrosion. This attribute ensures longevity and minimizes maintenance costs, making it an economical choice for businesses.

Moreover, the system incorporates various safety features. Automatic shut-off valves and alarms are integrated into the design to prevent overpressure situations, protecting both equipment and personnel. These built-in safeguards align with industry standards and promote a culture of safety within organizations.

The Pressure Systems 9022 also emphasizes energy efficiency. By optimizing the pressure regulation process, it reduces energy consumption, contributing to lower operational costs and a smaller environmental footprint. This focus on sustainability resonates with modern businesses looking to enhance their eco-friendly practices.

In conclusion, Pressure Systems 9022 stands out as a versatile, efficient, and reliable solution for pressure management. Its advanced features, durability, and commitment to safety make it a valuable asset for any industrial operation. By adopting this technology, organizations can enhance performance while ensuring compliance with safety regulations, ultimately leading to a more productive and sustainable operation.