Ready the module for multi-point calibration by

C 00 FFFF 3 1 64 (9016)

executing the Configure & Start (‘00’) sub-command of

 

 

‘C’. This establishes all the channels to be affected, and

or

determines the total number of calibration points that

 

 

will be supplied (3 in this example) in later steps. It also

C 00 0FFF 3 1 64 (9021/9022)

starts module averaging for calibration (64 samples in

 

 

this example). The linear fit (1) is required.

 

 

Apply 1st calibration pressure to the module’s CAL or

 

 

RUN inputs. The zero (0.0) point is assumed in this case.

 

 

After applying zero pressure verify that this pressure is

[rFFFF0]

[r0FFF0]

measured correctly by the module.

 

 

When the data are stable, enter the Collect Data (‘01’)

C 01

1 0.0

sub-command of ‘C’ specifying this first calibration

 

 

point (1) with zero pressure (0.0).

 

 

Apply 2nd calibration pressure to the module’s CAL or

 

 

RUN inputs. A full-scale (+5 psi) point is assumed in

 

 

this case.

 

 

Verify that pressure reads correctly.

[rFFFF0]

[r0FFF0]

When the data are stable, enter another Collect Data

C 01

2 5.0

(‘01’) sub-command of ‘C’ specifying this second

 

 

calibration point (2) with 5.0 psi pressure.

 

 

Apply 3rd calibration pressure to the module’s CAL or

 

 

RUN inputs. A mid-scale negative (-2.5 psi) point is

 

 

assumed in this case. A vacuum pump is normally

 

 

required to achieve such a pressure with 903x calibrators.

 

 

Verify that measured pressure reads correctly.

[rFFFF0]

[r0FFF0]

When the data are stable, enter last Collect Data (‘01’)

C 01 3 -2.5

sub-command of ‘C’ for this point (3) with a negative

 

 

(-2.5 psi) pressure.

 

 

Now that data have been collected for every point

C 02

originally specified, calculate and apply the new

 

 

coefficient data with a Calculate and Apply (‘02’) sub-

 

 

command of ‘C’. This also restores the module to using its original averaging parameters that existed before the first ‘C’ command.

4 - 12www.PressureSystems.com

Page 122
Image 122
Pressure Systems 9022 user manual

9022 specifications

Pressure Systems 9022 is a state-of-the-art pressure management system designed for various industrial applications, offering unparalleled efficiency and reliability. This innovative system has become a benchmark in pressure technology, providing users with advanced control and monitoring features.

One of the main features of Pressure Systems 9022 is its ability to deliver precise pressure regulation. Utilizing advanced algorithms, the system continuously monitors pressure levels and adjusts them in real-time, ensuring optimal performance and safety. This feature is particularly beneficial in industries such as oil and gas, pharmaceuticals, and food processing, where maintaining specific pressure levels is crucial for product integrity and process efficiency.

The 9022 is also equipped with cutting-edge connectivity options. With built-in IoT capabilities, the system can be easily integrated into existing networks, allowing for remote monitoring and control. Users can access pressure data in real-time through a user-friendly interface, which enhances decision-making and reduces response times in case of irregular conditions.

Another standout characteristic is its robust construction. Designed to endure harsh environmental conditions, the Pressure Systems 9022 is built from high-grade materials that provide excellent resistance to wear and corrosion. This attribute ensures longevity and minimizes maintenance costs, making it an economical choice for businesses.

Moreover, the system incorporates various safety features. Automatic shut-off valves and alarms are integrated into the design to prevent overpressure situations, protecting both equipment and personnel. These built-in safeguards align with industry standards and promote a culture of safety within organizations.

The Pressure Systems 9022 also emphasizes energy efficiency. By optimizing the pressure regulation process, it reduces energy consumption, contributing to lower operational costs and a smaller environmental footprint. This focus on sustainability resonates with modern businesses looking to enhance their eco-friendly practices.

In conclusion, Pressure Systems 9022 stands out as a versatile, efficient, and reliable solution for pressure management. Its advanced features, durability, and commitment to safety make it a valuable asset for any industrial operation. By adopting this technology, organizations can enhance performance while ensuring compliance with safety regulations, ultimately leading to a more productive and sustainable operation.